How does loading and sequencing output differ between the AVITI™ and AVITI24™ systems?
Question
How does loading and sequencing output differ between the AVITI and AVITI24 systems?
Answer
AVITI24 with software version 3.4.0 or later enables new computational improvements that enhance the performance of sequencing runs and increase pass-filter reads by up to 50% for many sequencing applications.
See the Cloudbreak™ Sequencing User Guide for library loading recommendations by assay for the AVITI and AVITI24 systems.
Summary
- Depending on the application, some libraries benefit from a slight loading increase to reach maximal output (up to 1.5 billion pass-filter reads) on AVITI24.
- For libraries that have previously been optimized for ~1 billion pass-filter reads using 2 x 150 or 2 x 75 Cloudbreak and Cloudbreak Freestyle™ kits on AVITI, increase the loading concentration by ~25% as a starting point on AVITI24.
- AVITI24 supersedes the gains from Expert Mode HD with most kit configurations while still maintaining high sequencing quality. Expert Mode HD is still available for 2 x 300 configurations on AVITI24.
Considerations
How much library should I load with AVITI24 to maximize output?
See the Cloudbreak Sequencing User Guide for library loading recommendations by assay for the AVITI and AVITI24 systems.
AVITI24 delivers up to 50% increased sequencing output for 2 x 75 and 2 x 150 Cloudbreak, Cloudbreak Freestyle, Trinity™, and Trinity Freestyle™ kits when running with software version 3.4.0 or later. Loading concentrations might need to be adjusted depending on the sequencing kit used and sequencing yield currently obtained on the AVITI system.
For most 2 x 75 and 2 x 150 Cloudbreak and Cloudbreak Freestyle runs, use a starting loading concentration of ~1.25X your current loading concentration on AVITI. For example, if you typically load your library at 10 pM to yield ~1 billion pass-filter reads on AVITI, start with 12.5 pM (a 25% increase) on AVITI24. To hit the 1.5 billion pass-filter polony metric, target 1.9 billion raw polonies before filtering. Your specific library type might require additional optimization of loading concentration. If you have optimized your library output with Cloudbreak or Cloudbreak Freestyle on AVITI or AVITI24 software version < 3.4.0 to yield 1.1–1.2 billion pass-filter reads with standard run settings, you might not need to change your current loading concentration.
For Trinity kits, start with your current loading concentration without additional material. See the Trinity Sequencing User Guide for more details.
The recommended loading concentrations are a starting point and loading can be titrated upward if runs maintain good quality and Occupancy Score below the optimal range. See the Knowledge Base article on Occupancy Score for additional details. For more information on loading and quantification tips to help you maximize pass-filter reads from your sequencing runs, see the Knowledge Base article on loading concentration.
How does increased sequencing output on AVITI24 compare to performance on AVITI with and without Expert Mode HD?
AVITI24 supersedes the gains from Expert Mode HD with most kit configurations while still maintaining high sequencing quality. Expert Mode HD is still available for 2 x 300 configurations on AVITI24. For libraries that have previously benefited from the use of Expert Mode HD on AVITI, see the following example comparison of a single cell RNA-Seq (scRNA-seq) library. This library was prepared with the 10X Chromium Single Cell 3’ Gene Expression library kit using human peripheral blood mononuclear cells, then end-polished before sequencing on the AVITI system. The library was sequenced with 28+90 cycle read length on AVITI, with Expert Mode HD on AVITI, and with increased (1.25X) loading on AVITI24. See the Enhanced Output with AVITI24 Technical Note for additional details.
Figure 1. (A) Million pass-filter reads and (B) Quality (%Q30) of an scRNA-Seq library sequenced on AVITI and AVITI24 with various settings. 1X Loading refers to the loading required to reach approximately 1 billion pass-filter reads for this library. 1X Loading + HD uses the same library and loading concentration with Expert Mode HD enabled on the AVITI. 1.25X Loading uses the same library on the AVITI24 with 25% higher loading concentration.
Compared to a standard run on AVITI, Expert Mode HD provides ~20% more pass-filter reads for this scRNA-Seq library. AVITI24 provides ~54% more pass-filter reads than the standard AVITI sequencing run for this library (Figure 1A). AVITI24 also provides ~338 million more pass-filter reads than AVITI with Expert Mode HD (Figure 1A). AVITI24 maintains comparable quality across runs, showing a slight increase over AVITI for this library (Figure 1B). Secondary analysis of these sequencing runs with Cell Ranger showed highly comparable metrics between all sequencing runs (Table 1). All runs also exceeded the Cell Ranger QC thresholds.
Table 1. Cell Ranger Metrics for scRNA-seq library sequenced with AVITI, AVITI with Expert Mode HD (AVITI + HD) and AVITI24.
| 10X Cell Ranger Metric | AVITI | AVITI + HD | AVITI24 |
|---|---|---|---|
| Number of Reads | 926,660,702 | 1,113,415,537 | 1,394,237,989 |
| Valid Barcodes | 97.50% | 98.00% | 97.90% |
| Estimated Number of Cells | 8,374 | 8,373 | 8,382 |
| Mean Reads Per Cell | 110,659 | 132,977 | 166,337 |
| Reads Mapped Confidently to genome | 94.00% | 94.10% | 94.10% |
| Median Genes per Cell | 3,150 | 3,232 | 3,284 |
| Q30 in Cell Barcode Region | 93.50% | 96.20% | 96.50% |
| Q30 in UMI Region | 93.50% | 95.00% | 95.30% |
| Q30 in RNA Read | 93.70% | 93.40% | 95.40% |
Will I still experience output improvements on AVITI24 without increasing library loading concentration?
Whole-exome sequencing (WES) libraries are a common PCR-plus library type sequenced on AVITI. To demonstrate the output enhancement with AVITI24, the following example compares a WES library sequenced with 1X loading on AVITI, 1X loading on AVITI24, and 1.25X loading on AVITI24.
Figure 2. (A) Million pass-filter reads and (B) Quality (%Q30) of a WES library sequenced on AVITI and AVITI24 with different loading schemes (1X and 1.25X). 1X Loading refers to the loading required to reach approximately 1 billion pass-filter reads for this library. 1.25X Loading uses the same library on the AVITI24 with 25% higher loading concentration.
With the same loading concentration (and same WES library), there is a 52% increase in pass-filter reads on AVITI24 (Figure 2A). A slightly higher loading concentration of 1.25X yields even more pass-filter reads while still maintaining high quality above 90% Q30 (Figure 2B).
To see data from other assays sequenced with AVITI and AVITI24, see the Enhanced Output with AVITI24 Technical Note.
For more information, reach out to your Field Applications Scientist or Element Biosciences Support at support@elembio.com.