How does increased sequencing output on AVITI24™ compare to performance on AVITI with and without Expert Mode HD?

AVITI24 with OS v3.4.0 or later enables new computational improvements that enhance the performance of sequencing runs and increase pass-filter reads by up to 50% for many sequencing applications. Depending on the application, some libraries benefit from a slight loading increase to reach maximal output (up to 1.5 billion pass-filter reads) on AVITI24.

Loading recommendations for different libraries are provided in the Cloudbreak Sequencing User Guide for both AVITI™ and AVITI24, and serve as a useful starting point. For additional information, please refer to the Knowledge Base articles How much library should I load with AVITI24 to maximize output? and Will I still experience output improvements on AVITI24 without increasing library loading concentration?

Comparison of AVITI with Expert Mode HD and AVITI24: scRNA-Seq Libraries

AVITI24 supersedes the gains from Expert Mode HD with most kit configurations (still available for 2x300 as this was not changed) while still maintaining high AVITI quality. For libraries that have previously benefited from the use of Expert Mode HD on AVITI, we’ve provided an example comparison of a scRNA-Seq library sequenced with standard settings on AVITI, with Expert Mode HD on AVITI, and with 1.25x loading on AVITI24.

Figure 1. (A) Million pass-filter reads and (B) Quality (%Q30) of an scRNA-Seq library sequenced on AVITI and AVITI24 with various settings. 1X Loading refers to the loading required to reach approximately 1 billion pass-filter reads for this library. 1X Loading + HD uses the same library and loading concentration with Expert Mode HD enabled on the AVITI. 1.25X Loading uses the same library on the AVITI24 with 25% higher loading concentration.

Compared to a standard run on AVITI, Expert Mode HD provides ~20% more pass-filter reads for this scRNA-Seq library. AVITI24 provides ~338 million more pass-filter reads than AVITI with Expert Mode HD, or an increase of ~54% over the standard AVITI sequencing run for this library (Figure 1A). AVITI24 also maintains comparable quality, showing a slight increase over AVITI for this library (Figure 1B). Secondary analysis of these sequencing runs with Cell Ranger showed highly comparable metrics between all sequencing runs (Table 1). All runs also exceeded the Cell Ranger QC thresholds.

Table 1. Cell Ranger Metrics for scRNA-seq library sequenced with AVITI, AVITI with Expert Mode HD (AVITI+HD) and AVITI24.

To see data from other assays sequenced with AVITI and AVITI24, see the Enhanced Output with AVITI24 Technical Note.

For more information, reach out to your Field Applications Scientist or Element Biosciences Support at support@elembio.com.