AVITI™ OS 3.2.0 Release Notes

Please Note: We recommend users initiate the update during a period of instrument downtime. During the update process, the system may restart multiple times as part of the upgrade procedure.

For systems with slower network connections and upgrading from AOS 2.x, the AOS 3.2.0 upgrade may take between 1-2 hours to complete. We recommend users initiate the update during a period of instrument downtime. If the upgrade process does not complete within 3 hours, please contact ElemBio Support at support@elembio.com.

Summary of Changes

AVITI OS v3.2.0

For AVITI24™ instruments:

  • Enhancements to reduce the total Cytoprofiling run time.
  • Cytoprofiling runs now include a thumbnail image in ElemBio™ Cloud, providing a visual indicator of cell confluency and cell culture health.
  • Additional tools are available to optimize system performance and manage resources, guided by Element Biosciences Support.

General Bug Fixes:

  • The PMG threshold for Trinity™ runs has been adjusted for better performance
  • Slow transfer speeds when transferring to USB have been improved

Known Issues, v3.2.0:

  • If leak errors are triggered for both sides of the instrument, AVITI OS cannot unlock the waste module door.
  • If the AFIC board loses power, it will not restart automatically and requires a power cycle. In the event of a power interruption or loss, power cycle the instrument via the UI after power is restored.
  • The run can continue even if polony map generation fails, as AVITI OS cannot detect a particular failure scenario.