AVITI OS v2.4.1 and v2.4.0 Release Notes

    Summary of Changes

    AVITI OS v2.4.1 (Patch release to v2.4.0)

    • To address potential run failure if demultiplexing barcodes for index1 and index2 are different lengths.

    AVITI OS v2.4.0

    Comprehensive support for new sequencing kits:

    • AVITI™ 2x300 Sequencing Kit Cloudbreak™ High Output
    • AVITI™ 2x300 Sequencing Kit Cloudbreak™ Medium Output

    Support for Low-Diversity Amplicon Libraries:

    • When you set up a run for the Adept Workflow and a 2 x 300 sequencing kit, you can enable a Low-Diversity Amplicon Library option. Enabling this option requires high sample index plexity but reduces the amount of PhiX spike-in required to sequence 16S amplicons and other low-diversity libraries. The option therefore increases throughput and decreases cost.

    Support for the AVITI System LT:

    • Visual indicators differentiate between the full AVITI System and the AVITI System LT, including new logos, new home screen background colors, and new start-up screens.
    • AVITI OS limits sequencing kit support for the AVITI System LT to the following medium and low output kits:
      • AVITI™ 2x300 Sequencing Kit Cloudbreak™ Medium Output
      • AVITI™ 2x150 Sequencing Kit Cloudbreak™ Medium Output
      • AVITI™ 2x150 Sequencing Kit Cloudbreak™ Low Output
      • AVITI™ 2x75 Sequencing Kit Cloudbreak™ Medium Output
    • High Output Kits Add-On
    • The High Output Kits add-on appears on the full AVITI System and any AVITI System LT that upgrades to the full AVITI system. The add-on lets users select high-output sequencing kits in addition to medium and low- output kits when setting up a sequencing run.

    Sequencing kit output levels appear for planned runs:

    • When applicable, sequencing kit output levels appear for planned runs on the instrument.

    Sustaining improvements:

    • If one library pool fails to generate data for a run that uses individually addressable lanes, the software now notifies the user that data is unavailable for the failed library pool.
    • For High Elevation mode, run setup estimates priming times more precisely than before.
    • Instrument start-up includes additional error handling for rare nest hardware failures.
    • Updates improve the process for software and firmware upgrades.
    • The Notifications list now includes pagination.

    Known Issues:

    • If the waste module leak errors are triggered for both instrument sides, the user cannot unlock the waste module door.
    • Low diversity runs can cause the thumbnail to have a green or yellow hue; an upcoming release will address this issue.